Dosimetric measurements of Onyx embolization material for stereotactic radiosurgery.
- Authors
Roberts, Donald A; Balter, James M; Chaudhary, Neeraj; Gemmete, Joseph J; Pandey, Aditya S
- Abstract
Arteriovenous malformations are often treated with a combination of embolization and stereotactic radiosurgery. Concern has been expressed in the past regarding the dosimetric properties of materials used in embolization and the effects that the introduction of these materials into the brain may have on the quality of the radiosurgery plan. To quantify these effects, the authors have taken large volumes of Onyx 34 and Onyx 18 (ethylene-vinyl alcohol copolymer doped with tantalum) and measured the attenuation and interface effects of these embolization materials.
- Publication
Medical physics, 2012, Vol 39, Issue 11, p6672
- ISSN
0094-2405
- Publication type
Journal Article
- DOI
10.1118/1.4757918