Prediction of residual astigmatism in cataract surgery at different diameter zones using optical biometry measurement.

Authors

Chang, Yin-Hsi; Pu, Christy; Lin, Ken-Kuo; Lee, Jiahn-Shing; Hou, Chiun-Ho

Abstract

The studies for astigmatism prediction error at different diameters using optical biometry are scant. We investigated patients who underwent cataract surgery with monofocal, nontoric intraocular lens (IOL) from 2017 through 2019 in a medical center. Patients with prior refractive surgeries, corneal opacity, or surgical complications were excluded. Corneal astigmatism (CA) was measured using AL-Scan at 2.4- and 3.3-mm diameter zones and calculated using the Barrett toric calculator preoperatively and postoperatively. The mean absolute error and centroid prediction error for the two zones were computed using double-angle plots. In total, 101 eyes of 76 patients were analyzed. Mean patient age was 68.7 ± 9.3 years and mean preoperative CA power was 0.7 ± 0.5 D. The overall centroid prediction error a 3.3 mm (0.09 ± 0.58 D@25) was significantly lower than that at 2.4 mm (0.09 ± 0.68 D@87) on the X-axis (P = 0.003). The 3.3-mm measurement also had a lower centroid prediction error than the 2.4-mm did for eyes with against-the-rule (ATR) and oblique astigmatism (P = 0.024; 0.002 on X-axis, respectively). The 3.3-mm measurement provided a more accurate CA estimation than the 2.4-mm did, particularly for ATR astigmatism. Diameter zone and astigmatism type should be considered crucial to precise astigmatism calculation.

Subjects

OPTICAL measurements; CATARACT surgery; ASTIGMATISM; CORNEAL opacity; INTRAOCULAR lenses; DIAMETER; CENTROID

Publication

Scientific Reports, 2022, Vol 12, Issue 1, p1

ISSN

2045-2322

Publication type

Academic Journal

DOI

10.1038/s41598-022-08253-6

EBSCO Connect | Privacy policy | Terms of use | Copyright | Manage my cookies
Journals | Subjects | Sitemap
© 2025 EBSCO Industries, Inc. All rights reserved