We found a match
Your institution may have rights to this item. Sign in to continue.
- Title
Solus™ flexible laryngeal mask patency fault.
- Authors
Stacey, K.; Chaggar, R. S.
- Abstract
The article reports on two separate incidents of apparent manufacturing faults of the Solus flexible wire reinforced laryngeal mask airway from medical device maker Intersurgical. It is noted that one fault almost completely occluded the airway device mid-way along its length, while in the other incident the occlusion was less severe and was observed following a period of high airway pressures.
- Subjects
MANUFACTURING defects; LARYNGEAL masks; AIRWAY (Anatomy); INTERSURGICAL Inc.; MEDICAL equipment; RESPIRATORY obstructions; PRODUCT design; MEDICAL equipment reliability
- Publication
Anaesthesia, 2017, Vol 72, Issue 7, p918
- ISSN
0003-2409
- Publication type
letter
- DOI
10.1111/anae.13940