- Title
A Sequential Comparison on the Risk of Haemorrhage with Different Sizes of Biopsy Needles for Stereotactic Brain Biopsy.
- Authors
Yuen, Jason; Zhu, Cannon X.L.; Chan, Danny T.M.; Ng, Rebecca Y.T.; Nia, Wai; Poon, Wai Sang; Ng, Ho Keung; Mok, Vincent C.T.; Wong, Laurence K.S.; Cheung, Tom C.Y.; Siu, Deyond Y.W.
- Abstract
Aim: To compare the risk of postoperative haemorrhage with different sizes of brain biopsy needles. Patients and Method: A cohort of patients using a 2.5-mm outer diameter side-cutting biopsy needle was compared to a subsequent cohort using a 1.8-mm needle of the same type. All data were collected prospectively. A CT scan was done within 12 h after surgery. Any visible haemorrhage at the operated site was documented. Results: From 2007 to 2013, 54 stereotactic brain biopsies (all frameless except for one frame-based) were performed. The 2.5-mm group comprised 29 procedures from 2007 to 2009. The 1.8-mm group comprised the subsequent 25 procedures. The diagnostic yields were 90 and 96% in the 2.5- and the 1.8-mm group, respectively (p = 0.615). Comparing the 2.5- and the 1.8-mm group, haemorrhage was significantly reduced: incidence (72 vs. 40%, p = 0.016); size of haemorrhage (mean 7.2 vs. 2.6 mm, p = 0.002); proportion of haemorrhage size >10 mm (34.5 vs. 4%, p = 0.006). Symptomatic haemorrhage rates were 3.4 and 0.0% in the 2.5- and the 1.8-mm group, respectively (p = 1.00). Conclusion: The 1.8-mm outer diameter needle carried a lower risk of postoperative haemorrhage than the 2.5-mm one, without compromising the diagnostic yield. © 2014 S. Karger AG, Basel
- Publication
Stereotactic & Functional Neurosurgery, 2014, Vol 92, Issue 3, p160
- ISSN
1011-6125
- Publication type
Article
- DOI
10.1159/000360862