We found a match
Your institution may have rights to this item. Sign in to continue.
- Title
Failed TAVI in TAVI Implantation: TAVI Dislocation Followed by Ensuing Surgical Graft Resection.
- Authors
Novotný, Róbert; Hlubocký, Jaroslav; Kovárník, Tomáš; Mitáš, Petr; Hlubocka, Zuzana; Rulíšek, Jan; Gulmez, Sevim Ismihan; Ghotra, Shubjiwan Kaur; Lindner, Jaroslav
- Abstract
We are presenting a case report of failed valve-in-valve treatment of severe aortic stenosis. A control ultrasonography after TAVI implantation revealed a severe aortic regurgitation of the graft which was subsequently unresolved with postimplantation dilatation. Second TAVI was implanted with cranial dislocation to the aortic root. Patient underwent a CT examination to clarify the TAVI in TAVI position. Patient underwent a surgical resection of TAVI with implantation of biological aortic valve prosthesis. In situations where TAVI treatment fails or is complicated beyond the possibility of endovascular repair, surgical intervention despite its higher risks is the preferred choice.
- Subjects
AORTIC stenosis; AORTIC stenosis treatment; ULTRASONIC imaging; SURGICAL excision; AORTIC valve surgery; CATHETERIZATION; DIAGNOSIS
- Publication
Case Reports in Cardiology, 2017, p1
- ISSN
2090-6404
- Publication type
Article
- DOI
10.1155/2017/5086586