We found a match
Your institution may have rights to this item. Sign in to continue.
- Title
Vergleichende Untersuchung von Einwegtransducern zur invasiven Blutdruckmessung.
- Authors
Matzek, F.; Boenick, U.; Frucht, U.; Schroeder, P. -M.
- Abstract
This paper reports on the results of a study in which the offset drift, sensitivity error and drift, as well as the linearity error of six DPT's were investigated under normal ambient conditions and varying conditions of temperature, light, operating time and Variation of supply voltage. In addition, the dynamic response, and the influence of storage at extreme temperatures, and resterilisation, was examined. The electrical and mechanical safety/reliability of DPT's was also investigated. The results obtained are compared.
- Publication
Biomedical Engineering / Biomedizinische Technik, 1989, Vol 34, Issue 3, p34
- ISSN
0013-5585
- Publication type
Article
- DOI
10.1515/bmte.1989.34.3.34