High-throughput and sensitive analysis of urinary heterocyclic aromatic amines using isotope-dilution liquid chromatography-tandem mass spectrometry and robotic sample preparation system.
- Authors
Zhang, Li; Xia, Yang; Xia, Baoyun; Nicodemus, Keegan; McGuffey, James; McGahee, Ernest; Blount, Benjamin; Wang, Lanqing
- Abstract
Heterocyclic aromatic amines (HCAA) are listed by the US Food and Drug Administration (FDA) as harmful or potentially harmful constituents of tobacco smoke. However, quantifying HCAA exposure is challenging. In this study, we developed a sensitive, precise, and accurate isotope dilution, liquid chromatography-tandem mass spectrometry (LC-MS/MS) method to quantify urinary HCAAs in smokers and nonsmokers. The high-throughput robotic sample preparation system could handle a throughput of over 300 samples per day, while maintaining intra-day and inter-day imprecision and bias ≤10 %. The limits of detection of carcinogenic HCAAs ranged from 0.31 to 0.83 pg/mL. The validated method was applied to measure HCAAs in urine collected from smokers and non-smokers. This sensitive and efficient analytical method is ideal to support large-scale biomonitoring studies of HCAA exposure.
- Subjects
HETEROCYCLIC compounds; AROMATIC amines; AUTOMATION; HIGH throughput screening (Drug development); LIQUID chromatography; MASS spectrometry
- Publication
Analytical & Bioanalytical Chemistry, 2016, Vol 408, Issue 28, p8149
- ISSN
1618-2642
- Publication type
Article
- DOI