We found a match
Your institution may have rights to this item. Sign in to continue.
- Title
Pressure Correction in the Calibration of Heat Dissipation Sensors.
- Authors
Thomle, Jonathan N.; Zhang, Z. Fred; Dai, Gao L.
- Abstract
Core IdeasEffects of air pressure on heat dissipation sensors previously unknown are quantified.Effects of air pressure are repeatable and consistent, therefore they can be removed.Changes in heat dissipation sensor calibration reduces error and decreases time of procedure. Heat dissipation sensors (HDSs) infer soil matric potential based on the measured rate of temperature rise from a heat source. The process for calibrating these sensors may use a pressure plate apparatus (PPA) to remove water to a certain calibration point that is controlled by pressure. Conventionally, the pressure within the PPA is released before the temperature change calibration points are measured (ΔT0). This conventional HDS calibration approach can take several weeks. We propose that the equilibrated temperature change at pressure (ΔTp) can be corrected empirically to the value at zero pressure. Based on measurements of 14 HDSs between 0.02 and 0.1 MPa, ΔTp was on average 6.2% larger than ΔT0. After pressure correction, using an empirical equation, the difference between the corrected ΔTp and ΔT0 was only 0.28%. These results demonstrate that this empirical approach provides sufficient accuracy to achieve HDS calibration without performing a PPA pressure release.
- Publication
Vadose Zone Journal, 2019, Vol 18, Issue 1, p1
- ISSN
1539-1663
- Publication type
Article
- DOI
10.2136/vzj2019.05.0052