We found a match
Your institution may have rights to this item. Sign in to continue.
- Title
A Conductive Nanowire‐Mesh Biosensor for Ultrasensitive Detection of Serum C‐Reactive Protein in Melanoma.
- Authors
Lin, Zuan‐Tao; Li, Yaxi; Gu, Jianhua; Wang, Huie; Zhu, Zhuan; Hong, Xia; Zhang, Zijing; Lu, Qinqin; Qiu, Jingyi; Wang, Xifan; Bao, Jiming; Wu, Tianfu
- Abstract
Abstract: Detection of extracutaneous melanoma is still challenging and is of importance in improving survival rate. In this report, an ultrasensitive biosensor is constructed where a C‐reactive protein (CRP) aptamers based molecular recognition core and a conductive polypyrrole (PPy) nanowire mesh based signal amplifier are developed. The conductive PPy nanowire (less than 10 nm in diameter) mesh architecture is uniformly dispersed within polymeric matrix via template‐free in situ synthesis. Serum CRP levels are quantitatively analyzed through monitoring the conductance change caused by polymeric network shrinkage upon the aptamer‐CRP binding. The limit of detection (LOD) of the polymeric sensor for human CRP sample can reach 7.85 × 10−19 m. This CRP‐specific biosensor and a commercial CRP enzyme‐linked immunosorbent assay (ELISA) kit are used to perform side‐by‐side measurement of serum CRP in melanoma patients. The results indicate that this conductive polymeric senor is highly sensitive and selective in accurately discriminating melanoma patients from healthy controls using serum CRP as a biomarker, which is further validated by a commercial human CRP ELISA kit. Collectively, this novel ultrasensitive nanowire‐based polymeric biosensor may hold promise in biomarker detection and diagnosis of cancer.
- Subjects
NANOWIRES; BIOSENSORS; BLOOD proteins; MELANOMA; DETECTION limit; PATIENTS
- Publication
Advanced Functional Materials, 2018, Vol 28, Issue 31, p1
- ISSN
1616-301X
- Publication type
Article
- DOI
10.1002/adfm.201802482