Quality assurance and characterization of narrowband ultraviolet B devices for use at home: lessons from the HI‐Light Vitiligo Trial.
- Authors
Rogers, A.; Akram, P.; Batchelor, J.M.; Crutchley, J.; Grocki, M.; Haines, R.H.; Meakin, G.; O'Dowd, K.; Ravenscroft, J.; Thomas, K.S.
- Abstract
During the trial, quality control checks of all devices prior to issue to participants determined whether the spectral emissions remained the same as at characterization, whether device output was within our predetermined range and whether each device was electrically safe. The device irradiance was measured for 10 devices at the comb tip in the middle of the tube. Dear Editor, Handheld narrowband ultraviolet B (NB-UVB) devices are available for the treatment of various skin conditions, but their optical radiation omissions are often not thoroughly characterized.
- Subjects
VITILIGO; QUALITY assurance; ULTRAVIOLET spectrophotometry; ULTRAVIOLET radiation measurement
- Publication
British Journal of Dermatology, 2021, Vol 184, Issue 5, p954
- ISSN
0007-0963
- Publication type
Article
- DOI
10.1111/bjd.19630