Frequency doubling perimetry screening mode compared to the full-threshold mode.
- Authors
Stoutenbeek, Remco; Heeg, Govert P.; Jansonius, Nomdo M.
- Abstract
The diagnostic performance of the frequency doubling perimetry (FDT) C20-1 screening mode was compared to that of the C20 full-threshold mode. For the number of defectsp < 1% in the total deviation plot, both modes appeared to perform similarly in terms of sensitivity, specificity, and area under the receiver–operating characteristic (ROC) curve. Different cut-off points should be applied for both modes to obtain equal sensitivity and specificity values, and– related to that– for most subjects more defects were found in full-threshold mode than in screening mode. For the screening mode, we found a sensitivity of 0.91 and a specificity of 0.88 at a cut-off point of>0 defects, and an area under the ROC curve of 0.93.
- Subjects
GLAUCOMA; CHRONIC diseases; EYE diseases; OPHTHALMOLOGY; SENSITIVITY theory (Mathematics); DIAGNOSIS
- Publication
Ophthalmic & Physiological Optics, 2004, Vol 24, Issue 6, p493