Efficacy of low dose and ultra-low dose on the visibility of peri-implant fenestration and dehiscences: a computed tomography study.
- Authors
Aktuna-Belgin, Ceren; Serindere, Gozde; Belgin, Huseyin Berkay; Serindere, Mehmet; Orhan, Kaan
- Abstract
Purpose: This study aimed to evaluate the visibility of peri-implant fenestration and dehiscences on computed tomography (CT) images taken with 2 different doses. Material and methods: The defects were created on the apical of 6 implants randomly selected from 20 titanium implants placed in the ribs, and dehiscences were created on the cervical of 8 implants. No defects were created around 6 implants. Macroscopic analysis of the implanted ribs was accepted as the gold standard. From the samples, images were taken by using both ultra-low dose (80 kVp, 50 mA, 1.25 mm slice thickness) and low dose (100 kVp, 50 mA, 1.25 mm slice thickness) protocols in CT. The images obtained were evaluated using a 5-point scale. Results: No significant difference was found between the area under the receiver operating characteristic of ultra-low dose protocol and low dose protocol in both defects based on the Wilcoxon test (p > 0.05). Conclusions: The ultra-low dose protocol could be applied by adhering to the “as low as reasonably achievable” principle in the diagnosis of peri-implant defects.
- Subjects
COMPUTED tomography; RECEIVER operating characteristic curves
- Publication
Polish Journal of Radiology, 2022, Vol 87, Issue 1, pe24
- ISSN
1733-134X
- Publication type
Article
- DOI
10.5114/pjr.2022.112466