Time-effectiveness and convenience of transvaginal ultrasound probe disinfection using ultraviolet vs chlorine dioxide multistep wipe system: prospective survey study.
- Authors
Kyriacou, C.; Robinson, E.; Barcroft, J.; Parker, N.; Tuomey, M.; Stalder, C.; Gould, D.; Al‐Memar, M.; Bourne, T.; Al-Memar, M
- Abstract
<bold>Objectives: </bold>To compare the efficiency, ease of use and user satisfaction of two methods of transvaginal ultrasound probe high-level disinfection: ultraviolet-C radiation (UV-C) and a chlorine dioxide multistep wipe system.<bold>Methods: </bold>This was a prospective survey study. UV-C units were introduced into a busy early pregnancy assessment service and compared with a multiwipe system for disinfection. Before seeing each patient, healthcare professionals (HCPs) measured with a stopwatch the time taken to complete a cycle of disinfection using either UV-C or chlorine dioxide multistep wipes and responded to a quick-response (QR) code-linked survey. Additional essential tasks that could be completed before seeing the next patient during probe disinfection were also documented. Using another QR code-linked survey, data on ease of use, satisfaction with the system used and preferred system were collected. The ease of use and satisfaction with the system were rated on a 0 to 10 Likert scale (0 poor, 10 excellent). A free-text section for comments was then completed.<bold>Results: </bold>Disinfection using UV-C (n = 331) was 60% faster than the chlorine dioxide multiwipe system (n = 332) (101 vs 250 s; P < 0.0001). A greater number of tasks were completed during probe disinfection when using UV-C, saving a further 74 s per patient (P < 0.0001). The HCPs using UV-C (n = 71) reported greater ease of use (median Likert score, 10 vs 3; P < 0.0001) and satisfaction (median Likert score, 10 vs 2; P < 0.0001) compared with those using the multiwipe system (n = 43). HCPs reported that the chlorine dioxide system was time-consuming and environmentally unfriendly, while the UV-C system was efficient and easy to use. Overall, 98% of the HCPs preferred using the UV-C system.<bold>Conclusions: </bold>UV-C technology is more time-efficient and allows more essential tasks to be completed during disinfection. For a 4-h ultrasound list of 15 patients, the use of UV-C would save 55 min 45 s. HCPs found UV-C preferable and easier to use. © 2021 International Society of Ultrasound in Obstetrics and Gynecology.
- Subjects
TRANSVAGINAL ultrasonography; CHLORINE dioxide; LONGITUDINAL method; MEDICAL personnel; LIKERT scale; CHLORINE compounds; OXIDES; RESEARCH funding; STERILIZATION (Disinfection)
- Publication
Ultrasound in Obstetrics & Gynecology, 2022, Vol 60, Issue 1, p132
- ISSN
0960-7692
- Publication type
journal article
- DOI
10.1002/uog.24834