Göz İçi Basıncı ılçümünde Goldmann Applanasyon Tonometresi, Tonopen Avia, Oküler Cevap Analizörü Karşılaştırması ve Merkezi Kornea Kalınlığının ılçümlere Etkisi.
- Authors
Paşaoğlu, Işıl; Eren, Mümin Hakan; Demircan, Ali; Güngel, Hülya; Altan, Çiğdem; Paşaoğlu, Ercan
- Abstract
Purpose: To compare the intraocular pressure measurements obtained with a Goldmann applanation tonometer (GAT), Tono-Pen AVIA, and ocular response analyzer (ORA) and to investigate the effect of central corneal thickness (CCT). Materials and Methods: Seventy-seven eyes followed because of primary open angle glaucoma were studied prospectively. IOP was measured by GAT, Tono-Pen AVIA, and ORA. The patients were divided into 3 groups as follows: group 1, thin (<545 µüm); group 2, moderate (545-580 üm); and group 3, thick (>580 üm) according to CCT measured by ultrasonic pachymetry. Measurements were compared between the groups. Results: The average age was 57.31±10.87 years. The average CCT was 528.92±12.58 üm for group 1, 557.40±9.14 üm for group 2, and 601.90±19.01 üm for group 3. GAT, Tono-Pen AVIA, and ORA measurements were 15.6±3.3, 16.7±2.9, and 16.9±3.9 mmHg in the whole study group; 15.1±3.9, 16.2±3.0, and 16.9±4.4 mmHg in group 1; 15.4±2.8, 16.5±2.5, and 16.4±3.7 mmHg in group 2; and 16.4±3.1, 17.7±3.4, 17.6± 3.6 mmHg in group 3, respectively. GAT measurements were significantly lower than Tono-Pen AVIA and ORA measurements in the whole study group (p=0.001, p=0.000), but there was no significant difference between Tono-Pen AVIA and ORA (p=0.64). According to CCT, while GAT values were lower than Tono-Pen AVIA measurements in groups 2 and 3 (p=0.015, p=0.18), no significant difference was observed in group 1 (p=0.075). Moreover, GAT values were lower than ORA measurements in groups 1 and 2 (p=0.003 and p=0.020), but there was no significant difference in group 3 (p=0.068). Conclusion: IOP values measured with GAT were lower than those measured with ORA in thin corneas, and lower than those measured with Tono-Pen AVIA in thick corneas.
- Subjects
TONOMETRY; INTRAOCULAR pressure; THICKNESS measurement; COMPARATIVE studies; GLAUCOMA; CORNEA; TONOMETERS; ULTRASONIC imaging
- Publication
Journal of Glaucoma-Cataract / Glokom-Katarakt, 2012, Vol 7, Issue 2, p101
- ISSN
1305-9173
- Publication type
Article