The effect of steam sterilisation at 134 °C on light intensity provided by fibrelight Macintosh laryngoscopes.
- Authors
Bucx; Veldman; Beenhakker; Koster; Bucx
- Abstract
We studied the repeated effect of sterilisation on light intensity in laryngoscopes from Penlon, Riester, Heine (two different blades), Medicon and Upsher. Light intensity was measured by a light meter using two methods. Measurements were performed before the decontamination procedure was carried out and subsequently after each series of 25 procedures until a total of 200 cycles was reached. Using method 1 (and 2), the reduction in light intensity after 200 cycles was 100% (100%; no light emitted), 37% (13%), 75% (69%), 79% (60%), 37% (14%) and 63% (55%) for each blade, respectively.
- Subjects
STEAM as a disinfectant; LARYNGOSCOPES
- Publication
Anaesthesia, 1999, Vol 54, Issue 9, p875