Impact of a PMMA tube on performances of a Vereos PET/CT system adapted for BSL-3 environment according to the NEMA NU2-2012 standard.
- Authors
Kahlaoui, Nidhal; Naninck, Thibaut; Le Grand, Roger; Chapon, Catherine
- Abstract
Introduction: A Vereos PET/CT device was adapted to be compatible with the experimentation in large animals within BSL-3 environment. The aim of this study was to investigate the impact of this modification on the performance according to NEMA NU2-2012 standard. Methods: Spatial resolution, sensitivity, count rate performance, accuracies of corrections and image quality were assessed using the NEMA NU2-2012 standards before and after installation of a transparent poly-methyl methacrylate tube of 8 mm thickness, 680 mm diameter and 2800 mm long inside the tunnel of the system. In addition, CT performance tests were performed according to manufacturer standard procedure. Results: Although the presence of the tube led to a slight decrease in sensitivity, performance measurements were in accordance with manufacturer preconisation ranges and comparable to previous performance published data. Conclusion: Modifications of Vereos PET/CT system allowing its use in BSL-3 conditions did not affect significantly its performance according to NEMA NU2-2012 standard. Key points: Question. Does a BSL-3 compatible modification alter Philips Vereos PET/CT performances according to NEMA NU2-2012 standards? Pertinent findings. Our Vereos PET/CT system was modified by a wall separating BSL-1 and BSL-3 sides and an 8 mm thickness PMMA tube inserted into the bore of the camera in order to extend the BSL-3 containment along the bed movement. The performances of our modified system according to NEMA NU2-2012 standards were not significantly impacted by the modifications and were in accordance with the values prescribed by the manufacturer. Implications for patients care. Our clinical PET/CT device was modified for human infectious diseases studies in Non-Human Primates. This unusual set up may then provide truly transposable data from preclinical studies into clinical application in infected patients.
- Subjects
ANIMAL experimentation; TUBES; SPATIAL resolution; COMMUNICABLE diseases; NASOENTERAL tubes
- Publication
EJNMMI Physics, 2022, Vol 9, Issue 1, p1
- ISSN
2197-7364
- Publication type
Article
- DOI
10.1186/s40658-022-00450-6