Caracterização dosimétrica de feixes de fótons com detector de diamante.
- Authors
Sabino, Talita; Rodrigues, Laura Natal; Furnari, Laura; Watanabe, Érika Yumi; Menegussi, Gisela
- Abstract
Objective: The present study was aimed at characterizing a diamond detector and observing its behavior. Materials and Methods: The dosimetric characteristics of 6 MV photon beams were measured by utilizing an automatic PTW MP3 scanning system with a diamond detector, and compared with measurements performed with a PTW 31010 ion chamber. Results: Measurements of dose linearity, dose rate dependence and depth dose distribution as well as dose profiles demonstrated the dosimetric characteristics of the diamond detector. Conclusion: Some measurements with diamond detector and ion chamber were validated with results published in the literature, demonstrating a good behavior of the diamond detector as compared with the ionization chamber that is widely utilized for dosimetry in radiotherapy, indicating that the diamond detector is a good choice for small field dosimetry.
- Subjects
PHOTON beams; RADIATION dosimetry; DETECTORS; SCANNING systems; IONIZATION chambers; RADIOTHERAPY; DIAMONDS
- Publication
Radiologia Brasileira, 2012, Vol 45, Issue 1, p35