Inhibition of Bradycardia Pacing and Detection of Ventricular Fibrillation Due to Far-Field Atrial Sensing in a Triple Chamber Implantable Cardioverter Defibrillator.
- Authors
Vollmann, Dirk; Lüthje, Lars; Görtler, Georg; Unterberg, Christina
- Abstract
Oversensing of intracardiac signals or myopotentials may cause inappropriate ICD therapy. Reports on far-field sensing of atrial signals are rare, and inappropriate ICD therapy due to oversensing of atrial fibrillation has not yet been described. This report presents a patient with a triple chamber ICD and a history of His-bundle ablation who experienced asystolic ventricular pauses and inappropriate detection of ventricular fibrillation due to far-field oversensing of atrial fibrillation. Several factors contributed to the complication, which resolved after reduction of the ventricular sensitivity.
- Subjects
IMPLANTABLE cardioverter-defibrillators; BRADYCARDIA; CARDIAC pacing; MEDICAL equipment; THERAPEUTICS; HEART diseases
- Publication
Pacing & Clinical Electrophysiology, 2002, Vol 25, Issue 10, p1513