We found a match
Your institution may have access to this item. Find your institution then sign in to continue.
- Title
Calibration and Validation of Nine Objective Vision Screeners With Contact Lens-Induced Anisometropia.
- Authors
Arnold, Robert W.; Davis, Bethanne; Arnold, Laura E.; Rowe, Kayla S.; Davis, Jodi M.
- Abstract
Purpose: New objective photoscreeners are emerging with encouraging national guidelines and pediatric providers can be reimbursed for photoscreening. Methods: Nine objective screeners were applied to two emmetropic subjects with or without known power contact lenses to induce spherical and cylindrical anisometropia close to American Association for Pediatric Ophthalmology and Strabismus (AAPOS) thresholds. Results: The screeners produced near linear pupil crescents and estimated refractions for induced anisometropia. Most performed well with AAPOS 2003 validation; however, the iPhone (Apple, Cupertino, CA) was too sensitive, whereas the remote autorefractors using Vision in Preschoolers Study criteria were less sensitive. Conclusions: The new, interpreted objective screeners appear valid for detection of anismetropia.
- Subjects
VISION testing equipment; ANISOMETROPIA; CONTACT lens complications; EYE examination; VISUAL accommodation; EQUIPMENT &; supplies
- Publication
Journal of Pediatric Ophthalmology & Strabismus, 2013, Vol 50, Issue 3, p184
- ISSN
0191-3913
- Publication type
Article
- DOI
10.3928/01913913-20130402-02