Beam rate influence on dose distribution and fluence map in IMRT dynamic technique
- Authors
Ślosarek, Krzysztof; Grządziel, Aleksandra; Osewski, Wojciech; Dolla, Łukasz; Bekman, Barbara; Petrovic, Borislava
- Abstract
Abstract: Aim: To examine the impact of beam rate on dose distribution in IMRT plans and then to evaluate agreement of calculated and measured dose distributions for various beam rate values. Background: Accelerators used in radiotherapy utilize some beam rate modes which can shorten irradiation time and thus reduce ability of patient movement during a treatment session. This aspect should be considered in high conformal dynamic techniques. Materials and methods: Dose calculation was done for two different beam rates (100MU/min and 600MU/min) in an IMRT plan. For both, a comparison of Radiation Planning Index (RPI) and MU was conducted. Secondly, the comparison of optimal fluence maps and corresponding actual fluence maps was done. Next, actual fluence maps were measured and compared with the calculated ones. Gamma index was used for that assessment. Additionally, positions of each leaf of the MLC were controlled by home made software. Results: Dose distribution obtained for lower beam rates was slightly better than for higher beam rates in terms of target coverage and risk structure protection. Lower numbers of MUs were achieved in 100MU/min plans than in 600MU/min plans. Actual fluence maps converted from optimal ones demonstrated more similarity in 100MU/min plans. Better conformity of the measured maps to the calculated ones was obtained when a lower beam rate was applied. However, these differences were small. No correlation was found between quality of fluence map conversion and leaf motion accuracy. Conclusion: Execution of dynamic techniques is dependent on beam rate. However, these differences are minor. Analysis shows a slight superiority of a lower beam rate. It does not significantly affect treatment accuracy.
- Subjects
RADIOTHERAPY; ELECTRON accelerators; IRRADIATION; CANCER patients; CANCER treatment; COMPUTER software
- Publication
Reports of Practical Oncology & Radiotherapy, 2012, Vol 17, Issue 2, p97
- ISSN
1507-1367
- Publication type
Article
- DOI
10.1016/j.rpor.2012.01.004