Inappropriate Discharges of an Implantable Cardioverter Defibrillator Secondary to Automatic Adjustable Gain of Atrial Tachycardia.
- Authors
Schecter, Stuart O.; Greenberg, Steven M.; Hoch, David H.; Cesa, Michael; Levine, Joseph H.
- Abstract
Third generation implantable cardioverter defibrillators are capable of complex arrhythmia detection using sensing algorithms with automatic adjustable gain settings. We report a unique case where inappropriate sensing of atrial tachycardia in a patient with a His bundle ablation lead to satisfaction of ventricular fibrillation detection criteria.
- Subjects
IMPLANTABLE cardioverter-defibrillators; IMPLANTED cardiovascular instruments; TACHYCARDIA; ARRHYTHMIA; VENTRICULAR fibrillation; CARDIAC arrest
- Publication
Pacing & Clinical Electrophysiology, 1997, Vol 20, Issue 6, p1721