Failed Blade Implant After 25 Years in Function: Case Description and Histologic Analysis.
- Authors
Farronato, Davide; Fumagalli, Davide; Asa'ad, Farah; Pasini, Pietro M.; Mangano, Francesco; Rasperini, Giulio
- Abstract
This report describes a failed blade implant after 25 years in function in the left maxilla and the results of histologic analysis of peri-implant tissues. The blade was retrieved, and histologic findings revealed that it was not osseointegrated despite being successfully in function for so many years. The long-term function of blade implants without osseointegration can be attributed to biotolerance exhibited by the implant rather than actual osseointegration.
- Subjects
DENTAL implant complications; MAXILLA surgery; OSSEOINTEGRATED dental implants; OSSEOINTEGRATION; DENTAL implants -- Design &; construction; PREVENTION; EQUIPMENT &; supplies; BIOMEDICAL materials; DENTAL implants; COMPLICATIONS of prosthesis; MEDICAL device removal
- Publication
International Journal of Periodontics & Restorative Dentistry, 2018, Vol 38, Issue 2, pe29
- ISSN
0198-7569
- Publication type
Article
- DOI
10.11607/prd.3384