- Title
自动乳腺超声诊断系统测量乳腺癌大小 及预估乳腺癌T分期.
- Authors
袁 娜; 赵巧玲; 王斐倩; 贾钰丽; 阮骊韬
- Abstract
Objective To compare the feasibility of automated breast volume scanner automated breast ultrasound system (ABUS) and the traditional ultrasound (US) in measuring breast cancer size so as to evaluate their value in predicting breast cancer T staging. Methods We retrospectively recruited 60 women with breast cancer who had received US and ABUS. The maximal tumor diameter was measured as tumor size. Based on the actual postoperative tumor size in pathology, Bland-Altman analysis and intraclass correlation coefficient (ICC) were used to compare the values measured by US and ABUS. Then we made a preliminary study of the accuracy of US and ABUS in predicting breast cancer T staging. Results The best absolute agreement was shown between US and ABUS in measuring tumor size. Moreover, ABUS showed better agreement with histology than US [average difference (-1.09±3.61)mm vs. (-1.57±4.99)mm] with a higher ICC (0.93 vs. 0.86), especially for tumors which were more than 2cm. In addition, both US and ABUS could predict breast cancer T staging relatively accurately (82.1% vs. 87.5%). Conclusion Both US and ABUS showed good agreement with pathology in measuring tumor size. ABUS even outperformed US in assessing tumor size for tumors beyond 2 cm. Therefore, ABUS can be considered as an alternative to US in T staging of breast cancer.
- Publication
Journal of Xi'an Jiaotong University (Medical Sciences), 2019, Vol 40, Issue 3, p465
- ISSN
1671-8259
- Publication type
Article
- DOI
10.7652/jdyxb201903024