Halo artifacts of indwelling urinary catheter by inaccurate scatter correction in 18F-FDG PET/CT imaging: incidence, mechanism, and solutions.
- Authors
Magota, Keiichi; Numata, Naoto; Shinyama, Daiki; Katahata, Junya; Munakata, Yamato; Maniawski, Piotr J.; Kobayashi, Kentaro; Manabe, Osamu; Hirata, Kenji; Tateishi, Ukihide; Kudo, Kohsuke; Shiga, Tohru
- Abstract
Background: Halo artifacts from urinary catheters can occur due to inaccurate scatter correction, and the artifacts affect the tumor visibility in 18F-FDG PET/CT images. We investigated the incidence rate and the mechanisms of halo-artifact generation and explored several scatter correction techniques to prevent artifacts. Methods: We conducted patient and phantom studies. (1) We retrospectively reviewed the cases of patients who had undergone 18F-FDG PET/CT scans. To determine the frequency of halo-artifact generation, we used the patients' PET images with a standard scatter correction based on a tail-fitted single-scatter simulation (TF-SSS) using 4-mm voxel μ-maps (TFS 4-mm). (2) We performed phantom studies to evaluate the effects of a urine catheter and two scatter correction techniques, i.e., TF-SSS with 2-mm voxel μ-maps (TFS 2-mm) and a Monte Carlo-based single-scatter simulation (MC-SSS) using 4-mm voxel μ-maps (MCS 4-mm). The average standardized uptake values (SUVs) were measured for axial PET images. (3) Using the patients' data, we investigated whether TFS 2-mm and MCS 4-mm can eliminate the artifacts in the clinical images. Results: (1) There were 61 patients with urinary catheters; in five (8.2%), halo artifacts were observed in the TFS 4-mm PET images. (2) The phantom study clearly reproduced the halo artifacts in the TFS 4-mm PET images. The halo artifacts were generated when urine moved in the interval between the CT and PET imaging, and when the urinary catheter was placed in a circular shape. The SUVs for the TFS 4-mm and TFS-2mm PET images were underestimated at the halo-artifact regions, whereas the SUVs for the MCS 4-mm PET images were close to the true values. (3) The halo artifacts disappeared in the TFS 2-mm PET images in 4/5 patients but not 1/5 patient, whereas the halo artifacts were completely absent in the MCS 4-mm PET images in 5/5 patients. Conclusions: These data suggest that halo artifacts are caused if the PET images do not correspond to the physical material in the μ-maps, which induces the scatter correction error. With the MC-SSS, it was possible to accurately estimate the scatter without generating halo artifacts.
- Subjects
URINARY catheters; IMAGING phantoms; IMPLANTABLE catheters; FLUORODEOXYGLUCOSE F18; POSITRON emission tomography computed tomography; DIAGNOSTIC imaging
- Publication
EJNMMI Physics, 2020, Vol 7, Issue 1, p1
- ISSN
2197-7364
- Publication type
Article
- DOI
10.1186/s40658-020-00333-8