We found a match
Your institution may have access to this item. Find your institution then sign in to continue.
- Title
Phased-array combination for MR spectroscopic imaging using a water reference.
- Authors
Abdoli, Abas; Maudsley, Andrew A.
- Abstract
Purpose To evaluate methods for multichannel combination of three-dimensional MR spectroscopic imaging (MRSI) data with a focus on using information from a water-reference spectroscopic image. Methods Volumetric MRSI data were acquired for a phantom and for human brain using 8- and 32-channel detection. Acquisition included a water-reference dataset that was used to determine the weights for several multichannel combination methods. Results were compared using the signal-to-noise ratio (SNR) of the N-acetylaspartate resonance. Results Performance of all methods was very similar for the phantom study, with the whitened singular value decomposition (WSVD) and signal magnitude (S) weighting combination having a small advantage. For in vivo studies, the S weighting, SNR weighting and signal to noise squared (S/N2) weighting were the three best methods and performed similarly. Example spectra and SNR maps indicated that the SVD and WSVD methods tend to fail for voxels at the outer edges of the brain that include strong lipid signal contributions. Conclusion For data combination of MRSI data using water-reference information, the S/N2 weighting, SNR and S weighting were the best methods in terms of spectral quality SNR. These methods are also computationally efficient and easy to implement. Magn Reson Med 76:733-741, 2016. © 2015 Wiley Periodicals, Inc.
- Publication
Magnetic Resonance in Medicine, 2016, Vol 76, Issue 3, p733
- ISSN
0740-3194
- Publication type
Article
- DOI
10.1002/mrm.25992