We found a match
Your institution may have access to this item. Find your institution then sign in to continue.
- Title
Evaluierung des Analysensystems Abbott IMx™ unter besonderer Berücksichtigung der Analysenzeitermittlung.
- Authors
Neuwald, Christine; Bauer, K.; Kittl, Eva Maria; Sitte, Sonja; Fischer, G.
- Abstract
During evaluation period of the analyser system Abbott IMx™ a time study process according to the standards of the study-group for determination of analysis time of the German and Austrian society for Clinical Chemistry was performed. A novel technology called Microparticle capture Enzyme Immuno Assay (MEIA) is described. Results of precision, linearity, accuracy, carry over and stability of calibration curve and the protocol of the time study are presented. Results obtained and our practical experiences prove, thai the enzyme immunoassay analyzer evaluated works with high precision and low expenditure of work.
- Publication
Journal of Laboratory Medicine / Laboratoriums Medizin, 1990, Vol 14, Issue 2, p73
- ISSN
0342-3026
- Publication type
Article