Feasibility and reliability of an automated controller of inspired oxygen concentration during mechanical ventilation.
- Authors
Saihi, Kaouther; Richard, Jean-Christophe M; Gonin, Xavier; Krüger, Thomas; Dojat, Michel; Brochard, Laurent
- Abstract
<bold>Introduction: </bold>Hypoxemia and high fractions of inspired oxygen (FiO2) are concerns in critically ill patients. An automated FiO2 controller based on continuous oxygen saturation (SpO2) measurement was tested. Two different SpO2-FiO2 feedback open loops, designed to react differently based on the level of hypoxemia, were compared. The results of the FiO2 controller were also compared with a historical control group.<bold>Methods: </bold>The system measures SpO2, compares with a target range (92% to 96%), and proposes in real time FiO2 settings to maintain SpO2 within target. In 20 patients under mechanical ventilation, two different FiO2-SpO2 open loops were applied by a dedicated research nurse during 3 hours, each in random order. The times spent in and outside the target SpO2 values were measured. The results of the automatic controller were then compared with a retrospective control group of 30 ICU patients. SpO2-FiO2 values of the control group were collected over three different periods of 6 hours.<bold>Results: </bold>Time in the target range was higher than 95% with the controller. When the 20 patients were separated according to the median PaO2/FiO2 (160(133-176) mm Hg versus 239(201-285)), the loop with the highest slope was slightly better (P = 0.047) for the more-hypoxemic patients. Hyperoxemia and hypoxemia durations were significantly shorter with the controller compared with usual care: SpO2 target range was reached 90% versus 24%, 27% and 32% (P < .001) with the controller, compared with three historical control-group periods.<bold>Conclusion: </bold>A specific FiO2 controller is able to maintain SpO2 reliably within a predefined target range. Two different feedback loops can be used, depending on the initial PaO2/FiO2; with both, the automatic controller showed excellent performance when compared with usual care.
- Publication
Critical Care, 2014, Vol 18, Issue 1, pR35
- ISSN
1364-8535
- Publication type
journal article
- DOI
10.1186/cc13734