We found a match
Your institution may have access to this item. Find your institution then sign in to continue.
- Title
2种X线曝光模式下双下肢全长摄影中图像质量和辐射剂量的比较.
- Authors
刘新丽; 李天然; 雷振兴; 刘刚; 宾盼; 徐启明
- Abstract
Objective To compare two exposure modes on image quality and radiation dose when used for double lower extremities full -length photography so as to provide guidance for clinical application. Methods Totally 60 patients undergoing double lower extremities full-length photography were selected with sequential experiment and then divided into two groups randomly and equally, of whom, there were four ones with genu varum. In the two groups, one group applied the exposure mode of AEC, and the other involved in FIXED mode. Then the two groups were compared on image quality and radiation dose. Results The exposure radiation to the patient under FIXED mode was higher than that under AEC mode although the common patients had the image quality with no significant difference in the two groups, and FIXED mode resulted in higher image quality for the patients with genu varum than AEC mode. Conclusion Exposure modes have effects on image quality and exposure dose, and it's suggested that AEC mode be applied for common patients from the aspect of radiation protection while FIXED mode be used for the patients with genu varum.
- Subjects
X-ray equipment; X-ray imaging; RADIOGRAPHY equipment; RADIATION exposure; IMAGE quality analysis; STANDARDS
- Publication
Chinese Medical Equipment Journal, 2018, Vol 39, Issue 5, p68
- ISSN
1003-8868
- Publication type
Article
- DOI
10.7687/j.issn1003-8868.2018.05.068