We found a match
Your institution may have access to this item. Find your institution then sign in to continue.
- Title
Contactless Stethoscope Enabled by Radar Technology.
- Authors
Lenz, Isabella; Rong, Yu; Bliss, Daniel
- Abstract
Contactless vital sign measurement technologies have the potential to greatly improve patient experiences and practitioner safety while creating the opportunity for comfortable continuous monitoring. We introduce a contactless alternative for measuring human heart sounds. We leverage millimeter wave frequency-modulated continuous wave radar and multi-input multi-output beamforming techniques to capture fine skin vibrations that result from the cardiac movements that cause heart sounds. We discuss contact-based heart sound measurement techniques and directly compare the radar heart sound technique with these contact-based approaches. We present experimental cases to test the strengths and limitations of both the contact-based measurement techniques and the contactless radar measurement. We demonstrate that the radar measurement technique is a viable and potentially superior method for capturing human heart sounds in many practical settings.
- Subjects
CONTINUOUS wave radar; STETHOSCOPES; RADAR; HEART sounds; MILLIMETER waves; SOUND measurement
- Publication
Bioengineering (Basel), 2023, Vol 10, Issue 2, p169
- ISSN
2306-5354
- Publication type
Article
- DOI
10.3390/bioengineering10020169