TYPE TESTING OF LiF:Mg,Cu,P (TLD-100H) WHOLE-BODY DOSEMETERS FOR THE ASSESSMENT OF H<sub>p</sub>(10) AND H<sub>p</sub>(0.07).
- Authors
Pereira, J S; Pereira, M F; Rangel, S; Caldeira, M; Carvalhal, G; Santos, L M; Cardoso, J V; Alves, J G
- Abstract
In this work, the initial results of the type testing of the LiF:Mg,Cu,P (TLD-100H) whole-body personal dosemeters are presented. An assessment of reproducibility, linearity of the response, the residual signal as a function of the dose, energy and angular dependence of the response was performed. In general, the dosemeters show good reproducibility for different dose values and a linear behaviour for a range between 0.1 and 300 mSv. The detection limits obtained are lower than 50 μSv. The system presents a good energy and angular response for different radiation qualities.
- Subjects
CONFORMANCE testing; ENERGY security; DETECTION limit
- Publication
Radiation Protection Dosimetry, 2019, Vol 184, Issue 2, p216
- ISSN
0144-8420
- Publication type
Article
- DOI
10.1093/rpd/ncy202