We found a match
Your institution may have access to this item. Find your institution then sign in to continue.
- Title
Design and numerical evaluation of an axial partial-assist blood pump for Chinese and other heart failure patients.
- Authors
Guang-Mao Liu; Dong-Hai Jin; Jian-Ye Zhou; Xi-Hang Jiang; Han-Song Sun; Yan Zhang; Hai-Bo Chen; Sheng-Shou Hu; Xing-Min Gui; Liu, Guang-Mao; Jin, Dong-Hai; Zhou, Jian-Ye; Jiang, Xi-Hang; Sun, Han-Song; Zhang, Yan; Chen, Hai-Bo; Hu, Sheng-Shou; Gui, Xing-Min
- Abstract
A fully implantable axial left ventricular assist device LAP31 was developed for Chinese or other heart failure patients who need partial support. Based on the 5-Lpm total cardiac blood output of Chinese without heart failure disease, the design point of LAP31 was set to a flow rate of 3 Lpm with 100-mmHg pressure head. To achieve the required pressure head and good hemolytic performance, a structure that includes a spindly rotor hub and a diffuser with splitter and cantilevered main blades was developed. Computational fluid dynamics (CFD) was used to analyze the hydraulic and hemodynamic performance of LAP31. Then in vitro hydraulics experiments were conducted. The numerical simulation results show that LAP31 could generate a 1 to 8 Lpm flow rate with a 60.9 to 182.7 mmHg pressure head when the pump was rotating between 9,000 and 12,000 rpm. The average scalar shear stress of the blood pump was 21.7 Pa, and the average exposure time was 71.0 milliseconds. The mean hemolysis index of LAP31 obtained using Heuser's hemolysis model and Giersiepen's model was 0.220% and 3.89 × 10-5% respectively. After adding the splitter blades, the flow separation at the suction surface of the diffuser was reduced. The cantilever structure reduced the tangential velocity from 6.1 to 4.7-1.4 m/s within the blade gap by changing the blade gap from shroud to hub. Subsequently, the blood damage caused by shear stress was reduced. In conclusion, the hydraulic and hemolytic characteristics of the LAP31 are acceptable for partial support.
- Publication
International Journal of Artificial Organs, 2017, Vol 40, Issue 9, p489
- ISSN
0391-3988
- Publication type
journal article
- DOI
10.5301/ijao.5000616