We found a match
Your institution may have access to this item. Find your institution then sign in to continue.
- Title
Difference in Bicarbonate Calculation Algorithm by Manufacturer Alters Diagnosis of Acid–Base Disorders.
- Authors
Tanemoto, Masayuki; Kimura, Takahide; Yamada, Seiki; Yokoyama, Takeshi; Okazaki, Yukio
- Abstract
The [HCO SB 3 sb SP - sp ] calculated by ALGO1 ([HCO SB 3 sb SP - sp ] SB ALGO1 sb ) and that calculated by ALGO2 ([HCO SB 3 sb SP - sp ] SB ALGO2 sb ) ranged from 11.2 to 34.2 mmol l SP -1 sp and from 10.7 to 33.6 mmol l SP -1 sp , respectively. [HCO SB 3 sb SP - sp ] SB ALGO1 sb , [HCO SB 3 sb SP - sp ] calculated by ALGO1 (mmol l SP -1 sp ); [HCO SB 3 sb SP - sp ] SB ALGO2 sb , [HCO SB 3 sb SP - sp ] calculated by ALGO2 (mmol l SP -1 sp ) [HCO SB 3 sb SP - sp ] SB ALGO1 sb , [HCO SB 3 sb SP - sp ] calculated by ALGO1 (mmol l SP -1 sp ); [HCO SB 3 sb SP - sp ] SB ALGO2 sb , [HCO SB 3 sb SP - sp ] calculated by ALGO2 (mmol l SP -1 sp ).
- Subjects
ACID-base imbalances; BICARBONATE ions
- Publication
Annals of Biomedical Engineering, 2019, Vol 47, Issue 12, p2351
- ISSN
0090-6964
- Publication type
Letter
- DOI
10.1007/s10439-019-02340-y